Pneumatic Vibration Isolation System 1VIS95 is of
a single piece design. It has automatic leveling valves and an autonomous
air supply Reservoir. All the equipment (leveling valves, hoses) is concealed
within the body of the support leg. The design looks clean.
Vertical and horizontal isolation is done by pneumatic springs. They have
low resonant frequencies thanks to very low spring constants. Resonant
frequency of air-springs is almost independent of load mass. Air to the
isolators is supplied from a reservoir. Depending on the mass of the table
top, the pressure in pneumatic isolators may vary between 0.5 and 1.9kg/cm².
The Reservoir may simply be filled with a car pump. The Reservoir should
be pumped with air not constantly, but only if required. The air pressure
in the Reservoir must be about 2 kg/cm². 3 support legs have
an automatic leveling valve each. The fourth leg has no automatic leveling,
it is connected to the leveling mechanism of another leg. The working
surface may be adjusted to level within a range of 12 mm with
precision of ±1mm. This is done by adjustment knobs located
on 3 support legs.
Choose the height of 1VIS95 support legs depending on the height of the
honeycomb table top. The recommended elevation of the surface of the table
top is about 900 mm.
Modifications and Ordering 1VIS95.
Ordering
for double load system 1VIS295 follows the same rules.
We offer the optimal Vibration Isolation Systems for the optical table
of your choice.
Here are the formulas how to calculate the required dimensions of 1VIS95.
B=0.56·b; L=0.56·l; H=900-h
The standard dimension values of 1VIS95 are as follows. Of these,
select the ones closely matching your calculations.
B: 500mm (code 05), 650mm (code 065), 800mm
(code 08);
L: 800mm (code 08), 1300mm (code 13), 1900mm(code
19);
H: 550mm (code 55), 600mm (code 60), 650mm
(code 65), 700mm (code 70), 750mm (code 75).
Examples of resulting codes for ordering:

We can produce 1VIS95 for individual table tops. Please specify custom dimensions explicitly without using the dimension codes.
A Car pump is available upon request. |
|