11PH - Photodetector

Photodetector
 
Large Apertures
10 mm Ø for the Silicon sensors
3 Version
• Silicon: 350 - 1080 nm, up to 750 mW
• Silicon-UV: 210 - 1080 nm, up to 25 mW
• Germanium: 800 - 1650 nm, up to 500 mW
Choice of Attenuators
• OD0.3: 50 % Transmission (for 11PH100-SiUV only)
• OD1: 10 % Transmission
• OD2: 1 % Transmission
High Accuracy
The new 11PH100-Si-HA presents the lowest calibration uncertainty to date
Precise Calibration
Wavelength selection in 1 nm steps
Smart Interface
Containing all the calibration data
 

11PH series silicon photodetectors provide a generous aperture for laser beams up to 10 mm in diameter. The threaded aperture allows you to install filters, attenuators, fiber optic adapters or other optics that suit your specific needs. All of the detectors have a NIST traceable calibration and Personal wavelength correction™. This is the choice for typical low power CW lasers from 210 to 1650 nm. To extend the performance of the photo detectors to higher measurable powers you can select one of our attenuators. The transmission of the OD1 is 10% and the transmission of the OD2 is 1%. When bought together, the detector is calibrated with and without the attenuator.

Previous models are currently on sale

Specifications
Model 11PH100-Si-HA 11PH100-SiUV 11PH20-Ge
MAX AVERAGE POWER (ALONE / WITH OD2) 36 mW / 750 mW 4 mW / 30 mW 30 mW / 500 mW
EFFECTIVE APERTURE 10 mm Ø 10 mm Ø 5 mm Ø
MEASUREMENT CAPABILITY
Spectral Range 350 – 1080 nm 210 – 1080 nm 800 – 1650 nm
   With OD0.3 --- 210 – 1080 nm ---
   With OD1 420 – 1080 nm 400 – 1080 nm 900 – 1650 nm
   With OD2 630 – 1080 nm --- 950 – 1650 nm
Maximum Measurable Power a 36 mW @ 1064 nm 4 mW @ 532 nm 30 mW @ 1064 nm
   With OD0.3 --- 11 mW @ 300 nm ---
   With OD1 300 mW @ 1064 nm 25 mW @ 532 nm 300 mW @ 1064 nm
   With OD2 750 mW @ 1064 nm --- 500 mW @ 1064 nm
Noise Equivalent Power b 10 pW @ 980 nm 10 pW @ 850 nm 60 pW @ 1550 nm
Rise Time (nominal) 0.2 sec 0.2 sec 0.2 sec
Peak Sensitivity 0.5 A/W @ 980 nm 0.45 A/W @ 850 nm 0.98 A/W @ 1550 nm
Calibration Uncertainty ±6.0 % (350 - 399 nm)
±2.0 % (400 - 449 nm)
±1.5 % ( 450 - 940 nm)
±2.0 % (941 - 980 nm)
±5.0 % (981 - 1049 nm)
±7.0 % (1050 - 1080 nm)
±8 % (200 - 219 nm)
±6.5 % (220 - 399 nm)
±2.5 % (400 - 899 nm)
±3.5 % (900 - 999 nm)
±5 % (1000 - 1049 nm)
±7 % (1050 - 1080 nm)
±3.5 % (800 - 1650 nm)
---
---
---
---
---
Calibration Uncertainty (with OD filters) ±4.0 % (420 - 980 nm)
±5.0 % (981 - 1049 nm)
±7.0 % (1050 - 1080 nm)
±6.5 % (210 - 399 nm)
±5 % (400 - 1049 nm)
±7 % (1050 - 1080 nm)
±5 %
DAMAGE THRESHOLDS
Maximum Average Power Density 100 W/cm2 100 W/cm2 100 W/cm2
PHYSICAL CHARACTERISTICS
Effective Aperture 10 mm Ø 10 mm Ø 5 mm Ø
Dimensions 38.1O x 27.4D mm 38.1O x 27.4D mm 38.1O x 27.4D mm
Weight (head only) 130 g 130 g 130 g
Specifications are subject to change without notice

a. Maximum value depends on the monitor.
b. Nominal value. Depends on environmental electromagnetic interference and wavelength.

Compatible Monitors:


11MAESTRO

11UNO

11TUNER

11M-LINK

11P-Link
 
Ordering Information
Model
11PH100-Si-HA
11PH100-Si-HA-OD1
11PH100-Si-HA-OD2
11PH100-SiUV
11PH100-SiUV-OD0.3
11PH100-SiUV-OD1
11PH20-Ge
11PH20-Ge-OD1
11PH20-Ge-OD2
 


Photodetector