11QE-B - Laser Energy Detectors

Laser Energy Detectors
 
Very Low Noise Level
Noise levels of a photo detector, but with the high energies of a pyroelectric: 50 nJ with the MT coating, 100 nJ with the BL coating
2 Coatings Available
BL: Black coating, sensitivity of 900 V/J, readings up to 400 Hz
MT: Metallic coating, sensitivity of 2400 V/J, readings up to 1000 Hz
Smart Interface
Containing all the calibration data
 

11QE-B laser energy detectors can measure energy up to 3.6 J, max repetition frequency - 1000 Hz. Very low noise levels - just 50 nJ with the MT coating and 100 nJ with the BL coating. Effective aperture - 7.8 x 7.8 mm. These detectors are very compact and convenient to use. They are suitable for many OEM, manufacturing and laboratory applications.

Specifications
Detector 11QE8SP-B-BL 11QE8SP-B-MT
MAX MEASURABLE ENERGY 3.6 mJ 1.3 mJ
MAX REPETITION FREQUENCY 400 HZ 1 000 Hz
EFFECTIVE APERTURE 7.8 X 7.8 mm 7.8 X 7.8 mm
MEASUREMENT CAPABILITY
Spectral Range * 0.19 - 20 µm 0.19 - 20 µm
Max Measurable Energy
   With 11M-LINK 3.6 mJ 1.3 mJ
   With 11S-LINK 2.9 mJ 1.1 mJ
   With 11MAESTRO 2.5 mJ 0.93 mJ
Noise Equivalent Energy
   With 11M-LINK (with noise suppression)a 100 nJ (N/A) 50 nJ (3 nJ)
   With 11S-LINK 100 nJ 50 nJ
   With 11MAESTRO 150 nJ 80 nJ
Sensitivity 900 V/J 2400 V/J
Max Repetition Frequency 400 Hz 1000 Hz
Max Pulse Width 10 µs 10 µs
Rise Time (0-100%) 30 µs 30 µs
Calibration Uncertainty ± 4.0% ± 4.0%
Repeatability <0.5 % <0.5 %
DAMAGE THRESHOLDS
Maximum Average Power 0.5 W 0.5 W
Maximum Average Power Density
   1064 nm, 7 ns, 10 Hz 1 W/cm2 1 W/cm2
Maximum Energy Density
   1064 nm, 7 ns, 10 Hz 50 mJ/cm2 50 mJ/cm2
PHYSICAL CHARACTERISTICS
Effective Aperture 7.8 x 7.8 mm 7.8 x 7.8 mm
Sensor Pyroelectric Pyroelectric
Absorber BL MT
Connector DB-15 DB-15
Dimensions 38.1 O X 27.4D mm 38.1 O X 27.4D mm
Weight 91 g 91 g
Specifications are subject to change without notice

* For the calibrated spectral range, see the user manual.
a. Obtain a lower noise level using the Noise Suppression function in the PC 11M-LINK application. Setting the sampling size at 128 will get the lowest noise level possible.

Compatible Monitors:


11MAESTRO

11M-LINK

11S-LINK
 
Ordering Information
Model
11QE8SP-B-BL
11QE8SP-B-MT
 


Laser Energy Detectors