11UP19-VR - Laser Power Detector

Laser Power Detector
 
Modular Concept
Increase the power capability of your detector with 2 different cooling modules
Hight Peak Power Volume Absorber
• Perfect for high density beams
• Average power density of 700 W/cm2 prevents degradation caused by repetitive pulses
Compact Design
Only 21 mm thick (15S model)
Energy Mode
Measure single shot energy up to 40 J
Smart Interface
Containing all the calibration data
 
Specifications
Model 11UP19K-15S-VR 11UP19K-30H-VR
MAX AVERAGE POWER (CONTINUOUS / 1 MINUTE) 15 W / 20 W 30 W / 35 W
EFFECTIVE APERTURE 18 mm Ø 18 mm Ø
COOLING METHOD Convection Heatsink
MEASUREMENT CAPABILITY
Spectral Range *a 0.3 – 2.5 µm 0.3 – 2.5 µm
Noise Equivalent Power b 2 mW 2 mW
Rise Time (nominal) c 2.5 sec 2.5 sec
Sensitivity (typ into 100 k Ω load) d 0.34 mV/W 0.34 mV/W
Calibration Uncertainty e ±2.5 % ±2.5 %
Repeatability ±0.5 % ±0.5 %
Energy Mode
   Sensitivity 0.1 mV/J 0.1 mV/J
   Maximum Measurable Energy f 40 J 40 J
   Noise Equivalent Energy b 0.02 J 0.02 J
   Minimum Repetition Period 4.5 sec 4.5 sec
   Maximum Pulse Width 90 ms 90 ms
   Accuracy with energy calibration option ±5 % ±5 %
DAMAGE THRESHOLDS
Maximum Average Power Density g 700 W/cm2 700 W/cm2
   Pulsed Laser Damage Thresholds Max Energy Density Peak Power Density
   1064 nm, 360 µs, 10 Hz 40 J/cm2 111 kW/cm2
   1064 nm, 7 ns, 10 Hz 6 J/cm2 860 MW/cm2
   532 nm, 7 ns, 10 Hz 4 J/cm2 570 MW/cm2
   266 nm, 7 ns, 10 Hz 1 J/cm2 143 MW/cm2
PHYSICAL CHARACTERISTICS
Effective Aperture 18 mm Ø 18 mm Ø
Absorber (Volume Absorber) VR VR
Dimensions 50H x 50W x 20.6D mm 50H x 50W x 56.3D mm
Weight (head only) 0.16 kg 0.21 kg
Specifications are subject to change without notice

* For the calibrated spectral range, see the user manual.
a. Adjustment multipliers for wavelengths under 300 nm are not traceable.
b. Nominal value, actual value depends on electrical noise in the measurement system.
c. With STANDA 11MAESTRO, 11UNO, 11P-LINK, 11TUNER and 11S-LINK monitors.
d. Maximum output voltage = sensitivity x maximum power.
e. Including linearity with power.
f. For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).
g. At 1064 nm, 10 W CW.

Compatible Monitors:


11MAESTRO

11UNO

11TUNER

11M-LINK

11P-Link

11S-LINK
 
Ordering Information
Model
11UP19K-15S-VR
11UP19K-30H-VR
 


Laser Power Detector